citation |
- title:
- Ground based in situ observations of ozone at Mount Chacaltaya (BO0001R) using uv_abs
- dates:
- fileIdentifier:
- EBAS/BO0001R.20120101000000.20190125000000.uv_abs.ozone.air.6y.1h.BO01L_ThermoSci49i.BO01L_O3_CHC.lev2.nc
(CodeSpace: )
- citedResponsibleParty:
- see section contact (role: distributor)
- presentationForm:
- otherCitationDetails:
- {
"Data definition": "EBAS_1.1",
"Set type code": "TU",
"Timezone": "UTC",
"File name": "BO0001R.20120101000000.20190125000000.uv_abs.ozone.air.6y.1h.BO01L_ThermoSci49i.BO01L_O3_CHC.lev2.nc",
"File creation": "20190513091756",
"Startdate": "20120101000000",
"Revision date": "20190125000000",
"Version": "1",
"Version description": "Zero check values, wrong values, and outliers>mean+2stddev removed. Not adjusted for zero drift. Adjusted to traceable standard. ppb = nmol/mol. Reported at measured conditions (Pamb=543.7 hPa,BenchT=37.0 ºC)",
"Data level": "2",
"Period code": "6y",
"Resolution code": "1h",
"Sample duration": "1h",
"Orig. time res.": "1mn",
"Station code": "BO0001R",
"Platform code": "BO0001S",
"Station name": "Mount Chacaltaya",
"Station latitude": "-16.2000007629",
"Station longitude": "-68.0999984741",
"Station altitude": "5320.0 m",
"Measurement altitude": "5246.0 m",
"Measurement height": "6.0 m",
"Regime": "IMG",
"Component": "ozone",
"Matrix": "air",
"Laboratory code": "BO01L",
"Instrument type": "uv_abs",
"Instrument name": "ThermoSci49i",
"Instrument manufacturer": "Thermo",
"Instrument model": "49i",
"Instrument serial number": "CM11200038 (model code 49I-A3NAC)",
"Method ref": "BO01L_O3_CHC",
"Standard method": "SOP=GAW_209(2013)",
"Inlet type": "Hat or hood",
"Inlet description": "Downward-facing tube protected by custom-made plastic hood (cut PET bottle). 6m above the ground, close to the window",
"Inlet tube material": "Teflon",
"Inlet tube outer diameter": "6.35 mm",
"Inlet tube inner diameter": "5.8 mm",
"Inlet tube length": "1.2 m",
"Maintenance description": "Pressure checked every 3 months, inlet filter replaced every 3 months, ozone scrubber replaced if needed. No ozone generator, therefore no span checks.",
"Flow rate": "0.568 l/min",
"Zero/span check type": "automatic",
"Zero/span check interval": "1d",
"Humidity/temperature control": "None",
"QA1 measure ID": "WCC Empa Site Audit",
"QA1 date": "20171219",
"QA1 document URL": "https://www.wmo.int/pages/prog/arep/gaw/documents/RCC-III_2017.pdf",
"Organization": "BO01L, \"Atmospheric Physics Laboratory, ALP, UMSA, Campus Cota Cota calle 27, Endifico FCPN piso 3, , La Paz, Bolivia\", , , , , , , ",
"Frameworks": "ACTRIS GAW-WDCRG",
"Originator": "Andrade Flores, Marcos, mandrade@chacaltaya.edu.bo, Universidad Mayor de San Andres, UMSA, Laboratorio de Fisica de la Atmosfera - Instituto de Investigaciones Fisicas, Calle 27 Cota Cota, Campus Universitario, 00000, La Paz, Bolivia",
"Submitter": "Moreno Rivadeneira, Isabel, caisa.moreno@chacaltaya.edu.bo, Universidad Mayor de San Andres, UMSA, Laboratorio de Fisica de la Atmosfera - Instituto de Investigaciones Fisicas, Calle 27 Cota Cota, Campus Universitario, 00000, La Paz, Bolivia",
"Acknowledgement": "Request acknowledgement details from data originator"
}
|