citation |
- title:
- Ground based in situ observations of particle_number_size_distribution at Madrid (ES0021U) using smps
- dates:
- fileIdentifier:
- EBAS/ES0021U.20170101000000.20200528000000.smps.particle_number_size_distribution.aerosol.3y.1h.ES09L_TSI_3936_CIE1.ES09L_smps_01_107bins.lev2.nc
(CodeSpace: )
- citedResponsibleParty:
- see section contact (role: distributor)
- presentationForm:
- otherCitationDetails:
- {
"Data definition": "EBAS_1.1",
"Set type code": "TI",
"Timezone": "UTC",
"File name": "ES0021U.20170101000000.20200528000000.smps.particle_number_size_distribution.aerosol.3y.1h.ES09L_TSI_3936_CIE1.ES09L_smps_01_107bins.lev2.nc",
"File creation": "20200619023518",
"Startdate": "20170101000000",
"Revision date": "20200528000000",
"Version": "1",
"Version description": "initial revision",
"Data level": "2",
"Period code": "3y",
"Resolution code": "1h",
"Sample duration": "1h",
"Orig. time res.": "270s",
"Station code": "ES0021U",
"Platform code": "ES0021S",
"Station name": "Madrid",
"Station latitude": "40.456436",
"Station longitude": "-3.725631",
"Station altitude": "669.0 m",
"Measurement latitude": "40.456436",
"Measurement longitude": "-3.725631",
"Measurement altitude": "669.0 m",
"Measurement height": "3.0 m",
"Regime": "IMG",
"Component": "particle_number_size_distribution",
"Unit": "1/cm3",
"Matrix": "aerosol",
"Laboratory code": "ES09L",
"Instrument type": "smps",
"Instrument name": "TSI_3936_CIE1",
"Instrument manufacturer": "TSI",
"Instrument model": "3936",
"Instrument serial number": "70838465",
"Method ref": "ES09L_smps_01_107bins",
"Standard method": "SOP=Wiedensohler2012",
"Inlet type": "Impactor--direct",
"Inlet description": "The sampling line is a 3-m long, 1/4\" diameter pipe with a weatherproof inlet to prevent rain to enter.",
"Humidity/temperature control": "Nafion dryer",
"Humidity/temperature control description": "sample dried to below 40% RH with nafion dryer operated with reduced pressure sheath air",
"Volume std. temperature": "273.15 K",
"Volume std. pressure": "1013.25 hPa",
"Detection limit": "0.0 1/cm3",
"Detection limit expl.": "Determined only by instrument counting statistics, no detection limit flag used",
"Measurement uncertainty": "10.0 %",
"Measurement uncertainty expl.": "This instrument participated in an MPSS intercomparison campaign at TROPOS during November 2017 and passed the quality standards of ACTRIS and GAW. It showed a concentration 1% lower than the Reference Instrument using the TROPOS evaluation software",
"Zero/negative values code": "Zero possible",
"Zero/negative values": "Zero values may appear due to statistical variations at very low concentrations",
"QA1 measure ID": "WCCAP-MPSS-2017-6",
"QA1 date": "20171117",
"QA1 document URL": "http://www.actris-ecac.eu/files/ECAC-report-MPSS-2017-6-1.pdf",
"Organization": "ES09L, \"Centro de Investigaciones Energeticas Medioambientales y Tecnologicas, CIEMAT, Spain\", , , , , , , ",
"Frameworks": "ACTRIS",
"Originator": "Artinano, Begona, b.artinano@ciemat.es, Centro de Investigaciones Energeticas Medioambientales y Tecnologicas, CIEMAT, Environment Department, Av. Complutense 40, , E-28040, Madrid, Spain",
"Submitter": "Martin, Laura, laura.martin@ciemat.es, Centro de Investigaciones Energeticas Medioambientales y Tecnologicas, CIEMAT, Environment Department, Av. Complutense 40, , E-28040, Madrid, Spain",
"Acknowledgement": "Request acknowledgement details from data originator",
"Comment": "The system was checked for sample and sheath flows and high-voltage supply in a regular base. In addition, some latex calibrations were performed along the year to check the correct whole instrument operation."
}
|