citation |
- title:
- Ground based in situ observations of chemiluminescence_photolytic at Hyytiälä (FI0050R)
- dates:
- fileIdentifier:
- EBAS/FI0050R.20180101000000.20200430115501.chemiluminescence_photolytic..air.27mo.1h.FI03L_DTM_BLC_Thermo_42i-TL_SMR.FI03L_Mastgas.lev2.nc
(CodeSpace: )
- citedResponsibleParty:
- see section contact (role: distributor)
- presentationForm:
- otherCitationDetails:
- {
"Data definition": "EBAS_1.1",
"Set type code": "TI",
"Timezone": "UTC",
"File name": "FI0050R.20180101000000.20200430115501.chemiluminescence_photolytic..air.27mo.1h.FI03L_DTM_BLC_Thermo_42i-TL_SMR.FI03L_Mastgas.lev2.nc",
"File creation": "20201112211503",
"Startdate": "20180101000000",
"Revision date": "20200430115501",
"Version": "1",
"Version description": "Initial revision",
"Data level": "2",
"Period code": "27mo",
"Resolution code": "1h",
"Sample duration": "1h",
"Orig. time res.": "6mn",
"Station code": "FI0050R",
"Platform code": "FI0050S",
"Station name": "Hyytiälä",
"Station GAW-ID": "HYY",
"Station state/province": "Pirka",
"Station land use": "Forest",
"Station setting": "Rural",
"Station GAW type": "R",
"Station WMO region": "6",
"Station latitude": "61.85",
"Station longitude": "24.283333",
"Station altitude": "181.0 m",
"Measurement height": "34.0 m",
"Regime": "IMG",
"Matrix": "air",
"Laboratory code": "FI03L",
"Instrument type": "chemiluminescence_photolytic",
"Instrument name": "DTM_BLC_Thermo_42i-TL_SMR",
"Instrument serial number": "n/a+0335003694",
"Method ref": "FI03L_Mastgas",
"Standard method": "None",
"Inlet type": "Hat or hood",
"Inlet description": "Rain protected inlet with apertures in a machined piece of a PTFE rod; 157 m long sample line of 16/14 mm diam. PTFE tube; main flow 45 l/min; inlet + line heated at 5 C above air temp; 5 um PTFE filters at analyser inlet",
"Ozone correction": "Corrected for reaction with O3 in the inlet",
"Water vapor correction": "Corrected for water vapor quenching in CLD",
"Humidity/temperature control": "None",
"Humidity/temperature control description": "Passive, sample heated from atmospheric to lab temperature",
"Detection limit expl.": "Determined by taking 2 times precision (standard deviation) of instrument signal for zero air during calibration and divided by the square root of the number (10) of observations in 1 hour. No detection limit flag used",
"Measurement uncertainty expl.": "Calculated as expanded uncertainty (k=2). Includes reproducibility (signal precision) + systematic errors (accuracy): calibration gas accuracy, dilution accuracy and analyser span drift",
"Zero/negative values code": "Zero/negative possible",
"Zero/negative values": "Zero and negative values may appear due to statistical variations and calibration uncertainty below the detection limit",
"QA1 measure ID": "ACTRIS NO round robin 2012",
"QA1 date": "20120601",
"QA1 document name": "WP4- NA4: Trace gases networking: Volatile organic carbon and nitrogen oxides Deliverable D4.4: Results of Round Robin exercise of NO",
"QA1 document URL": "http://fp7.actris.eu/Portals/97/deliverables/PU/WP4_D4.4_M24.pdf",
"Organization": "FI03L, University of Helsinki, UHEL, Division of Atmospheric Sciences, PO BOX 64, , FI-00014, University of Helsinki, Finland",
"Originator": "Petäjä, Tuukka, tuukka.petaja@helsinki.fi, University of Helsinki, UHEL, \"Division of Atmospheric Sciences, Department of Physics\", PO BOX 64, , FI-00014, University of Helsinki, Finland",
"Submitter": "Keronen, Petri, , , , , , , , , ",
"Acknowledgement": "Request acknowledgement details from data originator"
}
|